Fluoride
LR Checker HC®, 0.00 to 2.00 ppm
With
the low-range fluoride colorimeter you achieve more precise technology
than your standard chemical test kit. You won’t have to struggle to
determine colors visually, our handheld colorimeter eliminates the hassle
of matching to a color chart by providing a direct digital fluoride
result.
The
HI-729 Checker® HC
(Handheld Colorimeter) provides a simple, accurate and cost-effective way
to measure fluoride.
Fluoride
is one of the few chemicals that have been shown to cause significant
effects in people through drinking water. Fluoride has beneficial effects
on teeth at low concentrations in drinking water, but excessive exposure
to fluoride in drinking water, or in combination with exposure to fluoride
from other sources, can give rise to a number of adverse effects.
Water
fluoridation is the controlled addition of fluoride to a public water
supply to reduce tooth decay. Fluoridated water operates on tooth
surfaces: in the mouth, it creates low levels of fluoride in saliva, which
reduces the rate at which tooth enamel demineralizes and increases the
rate at which it remineralizes in the early stages of cavities. A 1994
World Health Organization expert committee suggested a level of fluoride
from 0.5 to 1.0 mg/L, depending on climate. Bottled water typically has
unknown fluoride levels, and some domestic water filters remove some or
all fluoride.
The
HI-729 Checker® is
simple to use. For this measurement you need 2 vials (#1 and #2); 2
mL of reagent is added in Deionized water is then added to vial #1 while
the sample to be tested is added to vial #2. Both vials are then gently
inverted 5 times, and wait 2 minutes. Cuvette #1 is then inserted into the
HI-729 Checker and the button pressed to zero; after that, insert cuvette
#2 to take a measurement.
Features
-
Easier to use and more accurate than chemical test kits
-
Dedicated to a single parameter
-
Small size, big convenience
-
Ideal for drinking water, environmental, industrial
Specifications
Fluoride
Range |
0.00
to 2.00 ppm |
Fluoride
Resolution |
0.01
ppm |
Fluoride
Accuracy |
±0.10
ppm ±5% of reading |
Fluoride
Method |
adaptation
of the Standard Methods for the Examination of Water and Wastewater,
18th edition, SPADNS method. |
Photometer/Colorimeter
Light Source |
LED
@ 575 nm |
Photometer/Colorimeter
Light Detector |
silicon
photocell |
Battery
Type/Life |
(1)
1.5V AAA |
Automatic
Shut-Off |
after
ten minutes of non-use |
Environment |
0
to 50°C (32 to 122°F); RH max 95% non-condensing |
Weight |
64
g (2.25 oz.) |
Dimensions |
81.5mm
x 61mm x 37.5mm (3.2" x 2.4" x 1.5”) |
Product Information
Instruction Manual
Brand : Hanna Instruments